Skip to Main Content

Resonant column (RC) is a standard laboratory test for the dynamic characterization of soils. Mass polar moment of inertia (MPMI) of the drive plate of RC is required for the solution of equation of motion for a specimen-mass system in RC. In standard procedure, the MPMI is evaluated by testing of calibration probes. Studies have shown that torsional stiffness of the probes have significant effect on the evaluation of MPMI of drive plate due to contribution from base of RC. The extent of contribution increases with increase in torsional stiffness of probes and consequently error in MPMI increases. This study presents a new calibration method that does not use probes. The MPMI of the drive plate is evaluated by suspension of drive plate with polyamide wire of negligible torsional stiffness and damping. The modified equation of motion is used to evaluate the MPMI of drive plate. Results show that the variation of MPMI with frequency is very small compared to its variation from conventional procedure. The apparent values of MPMI from conventional procedure are found to be higher than its inferred true-value, which results in overestimation of shear wave velocity of soil specimen.

Resonant column (RC) is a laboratory standard (ASTM, 2007) for the evaluation of dynamic properties of geomaterials such as shear wave velocity (Vs). The mass polar moment of inertia (MPMI) of the drive plate (Io) is required for the solution of equation of motion in RC tests (e.g. Richart et al., 1970; Cascante et al., 2003). Io is typically evaluated by standard procedure in which calibration probes of different resonant frequencies (fo) and therefore different stiffness are tested.

In standard procedure, the evaluation of Io is not consistent even if the probes are made out of the same material. The discrepancy in Io is due to contribution from base of RC among other effects (Drnevich, 1978; Avramidis & Saxena, 1990; Khan et al., 2008; Clayton et al., 2009). The extent of contribution depends on the relative stiffness of the probe to that of the base. Amplitude of input excitation (current) and transfer function (TF) of peripheral instruments such as power amplifier can also affect the position of resonant frequency of the probes and therefore affect the Io (Khan et al., 2013).

In this paper, a new method (NM) to evaluate Io of drive plate is presented. In the proposed method drive plate is suspended with a thin polyamide wire and therefore the contribution from base of RC is absent. The response of suspended drive plate is governed by its inertia only because the damping coefficient (c) and stiffness (k) of suspension system is negligible. The validity of this assumption is verified by conducting free vibration of the drive plate and calculation of the torsional stiffness of the wire (Kw). The response of peripheral instruments does not affect the proposed method; however, amplitude of excitation has slight effect on the Io. The results show consistent values of Io at all values of frequencies within the tested bandwidth. The results also show that the apparent values of Io from conventional procedure are higher compared to its inferred true value and the difference increases as the stiffness of the calibration probes increases. Consequently, the Vs of the soil specimen is overestimated for Io values higher than its true value.

The specimen in an RC device under torsional excitation is modelled as a single-degree-of-freedom system (SDOF) if mass of the RC base is 100 times more than the mass of the specimen (Ashmawy & Drnevich, 1994) or MPMI of the base is 100 times larger than MPMI of the specimen (ASTM, 2007). Both conditions are difficult to fulfil in a typical RC device. Khan et al. (2013) showed that the base of the RC device shall not be assumed as fixed and that the apparent Vs shall be corrected depending on the ratio of torsional stiffness of the specimen (Ksp) to torsional stiffness of the base (Kb).

In standard procedure, the Io of drive plate is evaluated by testing calibration probes (fo = 20–100 Hz). The error in measured fo of probes and therefore in Io increases with increase in torsional stiffness of probes (e.g. Clayton et al., 2009; Sasanakul & Bay, 2010). Figure 1 illustrates the effect of fo on Io of drive plate from different studies. The curves in Fig. 1 are predictions from exponential models fitted to the original data to generate Io values at similar values of fo. The Io and fo values are then normalised to better visualise the trend. All studies indicate an increase of up to 45% in Io with fo ranging from 27 to 244 Hz. A similar percentage change (up to 20%) in Vs of a soil specimen is to be expected because Vs is proportional to Io. Since the Io values always increase with increase in stiffness of calibration probes; therefore, the true value of the Io shall correspond to calibration probe of zero stiffness. The conventional procedure will therefore always result in an overestimation of Vs depending on the chosen value of apparent Io. The effect of different values of Io on the Vs of a sand specimen (Barco 71) is presented in Fig. 2. The sand specimen is tested at an isotropic confinement of 32 kPa and at shear strains smaller than 10−6. The Io values in Fig. 2 vary from 0·008 to 0·0116 kg m2, which represent the largest change (45%) in Io reported in literature (Fig. 1). The Vs of the soil specimen consequently increases from 149  to about 180 m/s (increase of 21%). Figure 2 shows Vs values that are also corrected for the base effects in RC according to the procedure proposed by Khan et al. (2008). Considering the effect of fo of probes on the Io of drive plate and consequent error (over estimation) in Vs, an NM for the calculation of Io is developed and presented in the following.

Fig. 1.

Variation of Io predicted by exponential models fitted to original data. Normalised to fo = 27 Hz and minimum values of Io

Fig. 1.

Variation of Io predicted by exponential models fitted to original data. Normalised to fo = 27 Hz and minimum values of Io

Close modal
Fig. 2.

Effect of Io on shear wave velocity (Vs) for sand (Barco 71) specimen tested at confinement of 34 kPa and shear strain (γ) < 10−6

Fig. 2.

Effect of Io on shear wave velocity (Vs) for sand (Barco 71) specimen tested at confinement of 34 kPa and shear strain (γ) < 10−6

Close modal

The displacement response (x) of a specimen under torsional excitation in RC device is presented in equation (1) (e.g. Cascante et al., 2005).

1

where F is the induced force in the coils due to current, J ( = Io + Is/3) is the weighted mass polar moment of inertia (MPMI) of the specimen (Is) and driving plate (Io), c is the viscous damping coefficient, rm is the distance of magnets from the centre of the drive plate and ra is the radial distance of the accelerometer from the centre of the drive plate. Derivatives of acceleration and velocity are represented by dots on the variables. Lorentz equation (F = BIc) relates the magnetic-force factor (B) and current in coils (Ic) to yield force on the drive plate at a distance of rm. In the absence of a specimen, k and c can be ignored, and J reduces to (Io + ΔIm). Equation (1) becomes equation (2) after rearrangement and substitutions for F and /ra.

2

where α is the angular acceleration of the drive plate (/ra). Equation (2) is an equation of line with slope equal to Brm/(Io + ΔIm), dependent variable of α and independent variable of Ic. ΔIm is MPMI of a calibration mass added to the drive plate. B and rm are properties of coil–magnet system and do not change. The ratios of α and Ic are obtained by performing a sinusoidal chirp of frequencies ranging from about 30 to 200 Hz without and with addition of calibration mass, which results in slopes S1 and S2 as functions of frequency, respectively. Ideally, the slopes shall not change with frequency. The two slopes are then used for the calculation of Io from equation (3) as a function of frequency.

3

Figure 3 presents the experimental set-up for the suspension of drive plate in RC along with the layout of instruments. The drive plate is attached to a three-dimensional (3D)-printed holder for levelling and a wire for lifting. The MPMI of the 3D-printed holder (6·5 × 10−7 kg m2) is very small which is later subtracted from the Io. The other end of suspension wire is anchored to a point roughly 1 m above the drive plate. A signal generator, power amplifier, switch, filter amplifier, oscilloscope and spectrum analyser comprise of the main equipment set-up.

Fig. 3.

Layout of instruments and suspension mechanism of drive plate

Fig. 3.

Layout of instruments and suspension mechanism of drive plate

Close modal

The wire is a made from monofilament polyamide with diameter (dw) of about 0·5 mm. The shear modulus of the wire (Gw) is about 1150 MPa (manufacturer rating). The torsional stiffness (Kw) of the wire with length (l) is calculated from equation (4). The calculated Kw of the wire (7·1 × 10−6 N m) is very small compared to typical torsional stiffness of sand specimen at a confinement of 30 kPa in an RC test (around 1000 N m).

4

A sinusoidal chirp (30–200 Hz) is generated and amplified in power amplifier. The response of drive plate is measured with two accelerometers mounted at a distance of ra from the centre. The current in the coils (Ic) and voltage signals from accelerometers are filtered (low pass 500 Hz) and amplified (20 dB) in the filter amplifier. The time signals are monitored in oscilloscope whereas the TF (ratio of acceleration (voltage) to Ic) is processed in real-time spectrum analyser. A uniform spectral window and ten averages are used to obtain the TF in frequency domain.

The TF values are multiplied by a factor to convert the acceleration voltages to angular acceleration for compatibility with Equation (2). The conversion factor is a function of sensitivity of accelerometer (mV/g), acceleration due to gravity (g), radial distance (ra) and amplification (dB) if different from amplification of Ic signal. The converted TF values represent the slope (S1) in equation (2). The test is repeated after addition of mass of known ΔIm to the drive plate to obtain the slope S2. Finally, Io is computed from equation (3) as a function of frequency (30–200 Hz). TF approach is preferred due to its ability to eliminate noise due to spectral averaging and faster acquisition instead of repeating the test at different frequencies (alternate approach).

Figure 4 presents the free vibration decay of the drive plate, which is compared with the free vibration decay of a typical sand specimen from an earlier RC test. The drive plate is excited by a single frequency sinusoid. The switch (Fig. 3) stops the current to the coils and simultaneously triggers the oscilloscope for synchronous capture of free decay. The absence of plate vibration in Fig. 4(a) after termination of current confirms negligible stiffness of the wire and validates the assumption of neglecting k in equation (1).

Fig. 4.

Normalised free-vibration amplitudes at the driving plate for (a) suspended and (b) on sand specimen. Time is normalised with respect to the excitation period

Fig. 4.

Normalised free-vibration amplitudes at the driving plate for (a) suspended and (b) on sand specimen. Time is normalised with respect to the excitation period

Close modal

The amount of phase difference between the input (current) and the response (acceleration) is proportional to damping ratio of the system (Khan et al., 2011). Input and response in a system with zero damping ratio will be either in phase or out of phase by π radians depending on the TFs of the instruments in the circuit. Figure 4(a) supports the elimination of c from equation (1) due to phase difference of +π radians between the current and acceleration signals.

Typical TFs between the acceleration response (voltage) and current are compared in Fig. 5. The variation of TF values within frequency bandwidth of interest is less than 0·005%. The current to the coils shall be kept constant during the sinusoidal chirp; however, fluctuations can occur due to response of the power amplifier. Fluctuation in current induced by the power amplifier at different frequencies is monitored by isolating the power amplifier and measuring its TF. The effect of variations in current on the ability of proposed method to evaluate Io is insignificant. Figure 6 presents the TF of audio power amplifier used in this study. The proposed method performs very well in spite of using ordinary power amplifier which is not maintaining constant amplification of current at different frequencies.

Fig. 5.

Normalised amplitude of the TFs (acceleration/current) for the driving plate with and without the added mass. Less than 0·005% variation in the frequency band of interest

Fig. 5.

Normalised amplitude of the TFs (acceleration/current) for the driving plate with and without the added mass. Less than 0·005% variation in the frequency band of interest

Close modal
Fig. 6.

Normalised TFs of the audio power amplifier used in this study for high and low amplifications

Fig. 6.

Normalised TFs of the audio power amplifier used in this study for high and low amplifications

Close modal

Figure 7 presents the computed MPMI of the drive plate (Io) as a function of frequency. The Io values at two levels of average current in the coils (Ic) are compared. The average of the current is used to represent the two levels due to variations in current induced by power amplifier (Fig. 6). The upper level of current (0·2 A) in the coils represents high level setting, which is not typically used in calibration procedures. The effect is probably due to larger torsion in the drive plate and therefore higher contribution of Kw of the suspension wire. The change between smallest and highest values of Io with frequency (27–108 Hz) at typical power level (0·05 A) is about 2·25%, which is much smaller than change in Io values obtained from the probes (10–20% in this study). The change in Io is about 1% if average value of Io is considered. Figure 7 shows that Io values tend to converge with decrease in resonant frequencies (stiffness) indicating that the apparent values of Io from probes are higher than its true value.

Fig. 7.

Comparison of the MPMI computed using the suspended driving plate (Ic =  0·05 A and 0·20 A) and the probes in this study

Fig. 7.

Comparison of the MPMI computed using the suspended driving plate (Ic =  0·05 A and 0·20 A) and the probes in this study

Close modal

Alternatively, variation of angular accelerations (α) and current in the coils (Ic) for a single frequency can be plotted to obtain the desired slopes. The plot yields a straight line with high r2 values (>0·98) as shown in Fig. 8; however, this approach is not recommended. As noted in Fig. 7, the change in current level has an effect on Io. A larger variation in current for the computation of slopes will induce some error in Io.

Fig. 8.

Variation of angular acceleration (α) with excitation current in the coils (Ic) for the suspended driving plate with and without the added mass (trend lines shown)

Fig. 8.

Variation of angular acceleration (α) with excitation current in the coils (Ic) for the suspended driving plate with and without the added mass (trend lines shown)

Close modal

RC can also be used as torsional-shear device at very low frequencies if the value of magnetic force factor (B) is known. Typically, this value is obtained through indirect measurements. The proposed methodology provides direct estimate of the magnetic force factor from equation (2). The magnetic force factor (B) of the tested RC device is 16·2 N/A.

A new methodology is presented in this paper to evaluate the MPMI of the drive plate of an RC device. Unlike the standard method, which relies on probes to evaluate MPMI and consequently suffer from many testing biases, the NM is based on suspension of drive plate with a thin wire of practically negligible stiffness and damping. Following important conclusions can be drawn from the study.

  • The apparent values of Io from conventional procedure are higher than its inferred true value and the difference increases as the stiffness of the calibration probe increases.

  • The change in Io values at frequencies of interest in the proposed NM is about 2%, which is much smaller than change of 10–45% from conventional procedure involving probes.

  • The calculated Vs for a sand specimen increases (overestimation) by about 21% when Io increases by 45% from its inferred true value.

  • The TF approach in frequency domain is fast, eliminates noise in the signal by performing spectral averaging, and does not require individual measurements at different frequencies.

  • The magnetic-force factor of the coils can be reliably computed from the proposed method, which is needed if a RC device is to be used as torsional shear device.

  • The proposed method based on TF approach does not suffer from uncontrollable fluctuations of current from an ordinary audio power amplifier.

The authors acknowledge the support provided by Natural Sciences and Engineering Research Council of Canada (NSERC), the University of Waterloo, and in part, by the Open Access Program from the American University of Sharjah. The results and discussions are that of authors and do not reflect the opinion of institutions.

B

magnetic-force factor

c

viscous damping coefficient of the system

d

diameter of suspension wire

fo

resonant frequency of probes

Gw

shear modulus of suspension wire

Ic

current in the coils

Io

mass polar moment of inertia (MPMI) of drive plate

Kb

torsional stiffness of the base of resonant column

Kw

torsional stiffness of the suspension wire

k

stiffness of the system

l

length of suspension wire

ra

distance of accelerometers from the centre of drive plate

rm

distance of magnets from the centre of drive plate

Vs

shear wave velocity of soil specimen

α

angular acceleration of drive plate

ΔIm

MPMI of the additional mass

ASTM
(
2007
). D 4015-07: Standard test methods for modulus and damping of soils by the resonant-column method.
Annual Book of ASTM Standards, ASTM International
,
West Conshohocken, PA, USA
.
Ashmawy
,
A. K.
&
Drnevich
,
V. P.
(
1994
).
A general dynamic model for resonant column/quasi-static torsional shear apparatus
.
Geotech. Test. J.
17
, No.
3
,
337
348
.
Avramidis
,
A. S.
&
Saxena
,
S. K.
(
1990
).
The modified stiffened Drnevich resonant column apparatus
.
Jpn. Soc. Soil Mech. Found. Engng
30
, No.
3
,
53
68
.
Cascante
,
G.
,
Vanderkooy
,
J.
&
Chung
,
W.
(
2003
).
Difference between current and voltage measurement in resonant column testing
.
Can. Geotech. J.
40
, No.
4
,
806
820
.
Cascante
,
G.
,
Vanderkooy
,
J.
&
Chung
,
W.
(
2005
).
A new mathematical model for resonant-column measurements including eddy-current effects
.
Can. Geotech. J.
42
, No.
1
,
121
135
.
Clayton
,
C.
,
Priest
,
J.
,
Bui
,
M.
,
Zervos
,
A.
&
Kim
,
S.
(
2009
).
The Stokoe resonant column apparatus: effects of stiffness, mass, and specimen fixity
.
Géotechnique
59
, No.
5
,
429
437
.
Drnevich
,
V. P.
(
1978
).
Resonant-column problems and solutions
.
Dynamic Geotechnical Testing
ASTM STP 654
,
384
398
.
Khan
,
Z. H.
,
Cascante
,
G.
&
El Naggar
,
M. H.
(
2008
).
Linearity of the first torsional mode of vibration and base fixidity in resonant column
.
Geotech. Test. J.
31
, No.
1
,
587
606
.
Khan
,
Z. H.
,
El Naggar
,
M. H.
&
Cascante
,
G.
(
2011
).
Frequency dependent dynamic properties from resonant column and cyclic triaxial tests
.
J. Franklin Inst.
348
, No.
7
,
1363
1376
.
Khan
,
Z. H.
,
Cascante
,
G.
,
Moayerian
,
S.
&
Grabinsky
,
M.
(
2013
).
Evaluation of transfer function and dynamic properties from strain-controlled resonant column tests
.
Geotech. Test. J.
36
, No.
3
,
1
8
.
Richart
,
F. E.
,
Hall
,
J. R.
&
Woods
,
R. D.
(
1970
).
Vibrations of soils and foundations
.
Englewood Cliffs, NJ, USA
:
Prentice-Hall
,
414
.
Sasanakul
,
I.
&
Bay
,
J.
(
2010
).
Calibration of equipment damping in a resonant column and torsional shear testing device
.
Geotech. Test. J.
33
, No.
5
,
363
374
, .
This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

or Create an Account

Close Modal
Close Modal

Gift article access

As a benefit of your subscription, you can share temporary access to restricted articles.

Each link will stop working after 30 days or 10 uses. You may create up to 10 links in a 30 day period.

Please sign in to your personal account to gift article access.

Register

Gift article access

As a benefit of your subscription, you can share temporary access to restricted articles.

Each link will stop working after 30 days or 10 uses. You may create up to 10 links in a 30 day period.

Gift articles remaining: --

Gift article access

Each link will stop working after 30 days or 10 uses. You may create up to 10 links in a 30 day period.

Gift articles remaining: --

Gift article access

As a benefit of your subscription, you can share temporary access to restricted articles.

Each link will stop working after 30 days or 10 uses.

You have reached the limit of 10 links within a 30 day period.