Skip to Main Content

Adrenaline and hydrogen peroxide (H2O2) have neuromodulatory functions in the brain, and peroxide is also formed during reaction of adrenaline. Considerable interest exists in developing electrochemical sensors that can detect their levels in vivo due to their important biochemical roles. Challenges associated with the electrochemical detection of hydrogen peroxide and adrenaline are that the oxidation of these molecules usually requires highly oxidising potentials (beyond 1.4 V against silver (Ag)/silver chloride (AgCl)) where electrode damage and biofouling are likely and the signals of adrenaline, hydrogen peroxide and adenosine overlap on most electrode materials. To address these issues, the authors fabricated pyrolysed carbon (C) electrodes coated with oxidised carbon nanotubes. Using these electrodes for fast-scan cyclic voltammetric measurements showed that the electrode offers reduced overpotentials compared with graphite and improved resistance to biofouling. Adrenaline oxidises on this electrode at 0.75 (±0.1) V and reduces back at −0.2 (±0.1) V, while hydrogen peroxide oxidation is detected at 0.85 (±0.1) V on this electrode. The electrodes are highly sensitive with a sensitivity of 16 nA/μM for adrenaline and 11 nA/μM for hydrogen peroxide on an 80 μm2 electrode. They are also suitable for distinguishing between adrenaline, hydrogen peroxide and adenosine. Thus, these probes can be used for multimodal detection of analytes.

Fast-scan cyclic voltammetry (FSCV), where the voltage sweep is extremely fast, has received considerable attention for the detection of neurotransmitters and neuromodulators, as it increases sensitivity and measurement speed.1,2 The ability of FSCV to detect nanomolar concentrations of a target analyte using chemically modified electrodes3 is of particular interest to neuroscientists, as it allows the detection of physiologically relevant concentrations of various bioactive substances in the living brain.2 Challenges associated with the development of chemically modified electrodes as brain implants are the degradation of chemical probes over time, loss of sensitivity, overlapping signals from analytes, biofouling and immune responses at the site of implantation.2 These issues have limited the application of carbon (C) fibre electrodes, which are considered to be the gold standard material for in vivo and in vitro preclinical applications because other conventional materials have even less favourable properties.2 Moreover, carbon fibre electrodes commonly cannot resolve dopamine and serotonin,4 hydrogen peroxide (H2O2) and adenosine from one another,2,5,6 which limits their application in biochemically complex environments such as the cerebrospinal fluid.

Considerable interest exists in using carbon nanotubes (CNTs) as an alternative electron carrier. Thanks to their small size, faster electron-transfer kinetics, reduced overpotentials, biostability and resistance to biofouling, CNTs have become an attractive material for the electrochemical detection of neurotransmitters and neuromodulators.2 Due to their hydrophobic nature and π-bonding, they form aggregates when dispersed in most solvents.7 To address these issues, the functionalisation of CNTs can be performed with oxidising acids, which opens the ends of the CNTs and grafts oxygen (O) atoms onto the exposed edge planes, making them polar enough to disperse in alcohols and even water.7 This functionalisation also allows CNTs to act as a local proton donor, thereby reducing overpotential for redox reactions in which proton transfer is required.7 CNT-based surfaces have been used in electrochemistry for electrochemical detection of neurotransmitters and neuromodulators.8–12 

Neuromodulators are an important class of compounds that manipulate neuronal activity – for example, by altering the firing pattern of neurons.1 Adrenaline is a monoamine neurotransmitter that serves a dual purpose, both as a hormone and as a neurotransmitter.13 In the brain, adrenaline is responsible for various cognitive functions, including alertness and flight-or-fight response.14 Dysregulation of adrenaline is known to cause depression and anxiety.14 

Hydrogen peroxide, superoxide (•O2) and hydroxyl radicals (•OH) are generated as the end product of energy metabolism and are usually considered to be waste/toxic materials to cells. Recent evidence suggests that peroxide molecules also play a significant role in cellular signalling similar to neurotransmission.15–17 The dual nature of hydrogen peroxide as a neuromodulator and in energy generation makes being able measure it attractive.

Electrochemical detection of adrenaline on carbon electrodes is reported at a highly anodic potential above +1.3 V.13,18 The use of such highly oxidising potential causes degradation of electrodes and generation of reactive oxygen species at the surface of the electrode, thereby reducing the sensitivity of electrodes.13 These challenges have limited the ability to measure adrenaline in the brain. Another issue with adrenaline detection in the brain is interference from analytes such as hydrogen peroxide and adenosine, which have overlapping signals.1,19,20 Thus, the occurrence of identical signals for multiple analytes requires electrode materials on which the signals from these analytes can be resolved.21 Studies have shown that CNTs have faster electron-transfer kinetics than graphite,12 are resistant to biofouling particularly when oxidised22 and are able to discriminate between these analytes, as the overpotentials for monoamines are considerably different from what they are on graphite.4 Biofouling is an issue with adrenaline, as it has a tendency to polymerise when oxidised to form melanin, an insoluble polymer that precipitates on the surface.23,24 The authors here therefore exploited the favorable properties of oxidised CNTs, by coating them onto pyrolytic carbon electrodes to investigate the electrochemical dynamics of adrenaline and hydrogen peroxide. The authors here also applied FSCV, as this has proved capable of measuring the concentrations of similar compounds at speeds relevant to in vivo application.25 

A number of previous studies26–31 have investigated similar materials for the electrochemical detection of adrenaline and similar compounds. In most cases, these have been aiming at in vitro measurements with the potential application being urine. This allows the use of large-area electrodes, soft electrodes and toxic components. Here the authors have focused on developing an ultramicroelectrode that is stiff enough to insert into tissue and stay away from potentially unsafe materials. Nasirizadeh et al.26 used a similar approach to that used here, but with a much larger electrode and a different polymer matrix, as where they wished to measure the concentration of ascorbate, the authors here wished to exclude it. Unlike previous groups, the authors here are specifically working towards implantable sub-micrometre electrodes for measuring neurotransmitters in the presence of ascorbate fast enough to detect single releases. This publication represents a step in that direction.

Hydrogen peroxide, hydrochloric (HCl) and sulfuric acids (H2SO4), adrenaline, 4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid, sodium chloride (NaCl), potassium chloride (KCl), sodium bicarbonate (NaHCO3), magnesium chloride (MgCl2), monosodium phosphate (NaH2PO4) and Nafion 1100W (5% in alcohols) were sourced from Sigma-Aldrich, Germany. Quartz capillaries (outer diameter (o.d.) 1 mm, inner diameter (i.d.) 0.5 mm, length 7.5 cm) were purchased from Sutter Instrument. The water used in the experiments was deionised in-house using a Millipore column and had a resistivity of 18 MΩ.

Quartz capillaries (o.d. 1.0 mm and i.d. 0.5 mm) were pulled using a Sutter P-2000 puller (Sutter Instrument). The pulling parameters were as follows: heating temperature 750°C, velocity 50 and DEL 127 (instrument parameters), which gave a tip diameter of 0.8–1.5 μm. The tip was broken shorter to reach a final diameter of 25–30 μm. Other micropipette pullers will have their own parameters, but the main difference from normal micropipettes is the melting point of quartz.

Pyrolysed carbon microelectrodes were fabricated by performing pyrolysis of propane in a nitrogen environment. Standard cooking propane was fed into the micropipette at the large end at atmospheric pressure plus 1 bar. The other end was bathed in a flow of nitrogen from a cylinder flowing out of a wide quartz tube at 50 ml/min and heated with a small butane/air torch. Quartz capillaries were heated for 1 min at the narrower end for carbonisation. Once the desired duration had been reached, the heat was moved towards the broader end. Upon completion of pyrolysis, the electrodes were allowed to cool under nitrogen flow. The micropipette was held at a dull orange heat, which is why quartz was required for this step.

CNTs were functionalised by dispersing them in concentrated hydrochloric acid and concentrated sulfuric acid at a ratio of 3:1 (v/v). The solution was stirred with a magnetic stirrer for 48 h. This served to open the ends of the tubes, remove any metal catalyst and graft carboxylic acid groups to the edges. Washing was carried out by decanting off the acid and repeated washing and decanting with distilled water until pH 7.0 was reached. After washing, the CNTs were allowed to air-dry for 24–36 h. A similar method has been shown to generate carboxylate and alcohol groups at CNTs by previous researchers,32 although the authors were unable to detect these groups directly. The treated CNTs wetted with water and were dispersible in ethanol; untreated CNTs were not, due to their low surface energy.

The oxidised CNTs were suspended in a solution of Nafion® diluted with ethanol and sonicated for 30 min prior to use. The suspension was stable for 6–8 weeks. To obtain a thin coating of oxidised CNTs onto the electrode, the pyrolytic carbon electrodes were dip-coated in this solution for 5 s, shaken off and allowed to dry. On larger glassy carbon electrodes for reference, this treatment changed them optically from shiny black to dull black and the water contact angle from 70° to totally spreading. The coating was just about visible in field emission gun scanning electron microscopy images. However, owing to the very small size of the CNTs and their sharp tips, the authors were unable to image them successfully. The process allowed the deposition of a sufficient amount of CNTs to modify the surface properties of the electrode. Successful coating could be verified by a change in electrochemical signal in a phosphate buffer solution, as the tips of the micropipettes are too small for the change to be seen easily by eye.

Electrochemical detection of adrenaline and hydrogen peroxide was carried in an electrophysiological set up that had ports for delivery and outflow of solution.31,33–36 To mimic the rapid change in concentrations observable in the brain, a flow cell was set up that delivered the target analyte as a 5 s bolus, followed by a pause of 10 s. A constant flow of buffer was maintained at 2 ml/min serving as a background. Electrochemical detection of analytes was carried out using the ‘dopamine waveform’,21 wherein the electrode was cycled from −0.4 to 1.3 V and cycled back to −0.4 V at 400 V/s in the presence of a silver (Ag) wire, coated in silver chloride (AgCl) in a potassium chloride solution (3.5 M) serving as a reference electrode (silver/ silver chloride). Data were collected and stored offline for analysis.

The electrochemical behaviour of adrenaline on CNT-coated electrodes was studied by introducing 1 μM of adrenaline to an electrode being voltage scanned. The electrode was exposed to ten consecutive cycles of adrenaline (ten bolus additions) while cycling the potential, initially at 400 V/s. Multiple oxidation peaks and one reduction peak were observed in the cyclic voltammograms at this scan rate. One oxidation peak (0.9 V) was associated with oxidation of adrenaline, and a second peak was seen at 1.1 V. Monoamines, such as dopamine and serotonin, are known to polymerise onto surfaces, usually from the partially oxidised intermediate. This generates a non-conductive surface layer, which on sensors leads to loss in sensitivity, although adrenaline is not as bad.37,38 Hence, the authors chose to increase the scan rate to 600 V/s to limit this side reaction. Background-subtracted voltammograms show an oxidation peak for adrenaline at 0.75 (±0.1) V, and the product is probably reduced back at −0.2 (±0.1) V, although this peak is difficult to resolve from the background (Figure 1(a)). This is a difference voltammogram with the baseline (Figure 1(b)) measured in plain buffer. A broad oxidation peak is seen, suggesting that the oxidation of adenosine on pyrolytic-CNT electrodes is a multistep oxidation process occurring on the surface of the electrode.

Figure 1

Oxidation of 1 μM adrenaline from −0.4 to 1.3 V at 600 V/s and 10 Hz; adrenaline is oxidised at 0.75 (±0.1) V and reduced at −0.2 (±0.1) V. (b) Background current. The oxidation peaks at 0.4 and 1.1 V and the reduction peak at 0.9 V are likely to be due to surface groups formed during oxidation of carbon.39 (c) The peak current is directly proportional to the scan rate with an r2 = 0.96 fit, showing that the current is dominated by adsorbed species. The electrode has a sensitivity of 16 nA/μM on about 80 μm2. (d) Theoretical prediction for adrenaline, using a diffusion constant of 6 × 10−6 m2/s. The rate of electron transfer was kept at 1 × 10−6 m2/s; other parameters are standard

Figure 1

Oxidation of 1 μM adrenaline from −0.4 to 1.3 V at 600 V/s and 10 Hz; adrenaline is oxidised at 0.75 (±0.1) V and reduced at −0.2 (±0.1) V. (b) Background current. The oxidation peaks at 0.4 and 1.1 V and the reduction peak at 0.9 V are likely to be due to surface groups formed during oxidation of carbon.39 (c) The peak current is directly proportional to the scan rate with an r2 = 0.96 fit, showing that the current is dominated by adsorbed species. The electrode has a sensitivity of 16 nA/μM on about 80 μm2. (d) Theoretical prediction for adrenaline, using a diffusion constant of 6 × 10−6 m2/s. The rate of electron transfer was kept at 1 × 10−6 m2/s; other parameters are standard

Close modal

The authors further investigated the behaviour of adrenaline on the electrode surface, by varying the scan speed and keeping the concentration of adrenaline constant. As the scan rate was varied, the peak current increased linearly, suggesting that adrenaline binds to the surface of the electrodes (Figure 1(c)). This type of pre-adsorption increases the sensitivity of detection by increasing the amount of analyte available to the electrode for a short time, thereby allowing the detection of the nanomolar concentration of the analyte of interest.21,40 To investigate if this waveform could be used for longer-term measurement, the electrode was exposed to ten consecutive cycles of adrenaline for 10 min. No changes in the voltammogram over this time were observed, suggesting that the CNT surface combined with the increased scan rate limits the polymerisation of adrenaline onto CNT electrodes, thus providing long-term stability. To understand the oxidation of adrenaline on CNT electrodes better, the authors modelled the rate of electron transfer, diffusion coefficient and chemical rate constant. Mathematical modelling and previous work on glassy carbon electrodes40 suggest that adrenaline is oxidised at 1.1 V and reduced at −0.15 V against silver/silver chloride (Figure 1(d)). Experimentally, oxidation is measured at 0.75 V. The difference between theoretical and experimental data can be accounted for by heteroatoms that are expected to be present on the surface of the electrode shifting the voltage. This has been suggested before by other researchers;27,28,30 even the effect of abrasion can cause a shift in voltage.41–43 The magnitude of the current is difficult to model due to the likelihood of adsorption, the high specific surface area of CNTs and the way that the quartz capillaries break increasing the surface area. Example images of tips are shown in Figures 2(c) and 2(d), revealing that the break is typically at an angle to the short axis of the capillary, resulting in a higher end area.

Figure 2

Oxidation and reduction properties of adrenaline using the FSCV protocol. (a) Two-dimensional hotspot, wherein current is represented on the Y-axis, while voltage is represented on the X-axis. The time axis is converted into a number of repetitions on the Z-axis. The hotspot shows the oxidation and reduction spots along with the steadiness of current across a number of trials. (b) Match between experimental and simulation data, showing that the experimental peaks are far broader. The experimental parameters were −0.4 to 1.3 V and cycled back to −0.4 V at 600 V/s repeated at 10 Hz. The simulated data parameters were diffusion constant D of 6 × 10−6 cm2/s, concentration of 1 μM and number of electrons transported (n) = 2; the rate of electron transport was set to 10−6 cm2/s, and the chemical rate was 10−4 cm2/s. (c, d) Examples of what the tips of the electrodes look like, showing how the end lies at an angle to the main axis

Figure 2

Oxidation and reduction properties of adrenaline using the FSCV protocol. (a) Two-dimensional hotspot, wherein current is represented on the Y-axis, while voltage is represented on the X-axis. The time axis is converted into a number of repetitions on the Z-axis. The hotspot shows the oxidation and reduction spots along with the steadiness of current across a number of trials. (b) Match between experimental and simulation data, showing that the experimental peaks are far broader. The experimental parameters were −0.4 to 1.3 V and cycled back to −0.4 V at 600 V/s repeated at 10 Hz. The simulated data parameters were diffusion constant D of 6 × 10−6 cm2/s, concentration of 1 μM and number of electrons transported (n) = 2; the rate of electron transport was set to 10−6 cm2/s, and the chemical rate was 10−4 cm2/s. (c, d) Examples of what the tips of the electrodes look like, showing how the end lies at an angle to the main axis

Close modal

Comparing these results with a few other recent publications (Table 1) detecting adrenaline, the specific sensitivity per unit area is comparable with them, several of which performed well in comparison with other results. As the limit of detection depends on the electronics as well as the chemistry and size of the electrode, it is difficult to compare these values fairly. The requirement for sensitivity is particularly acute in electrodes for implantation or scanning over a surface, as their surface area cannot be increased without impacting their function in other ways.

Table 1

Specific sensitivities of electrodes recently proposed in the literature for adrenaline sensing (approximate)

ReferenceSensitivity: (μA/mM)/mm2
Zare et al.31 0.2
Reddy et al.27 42
Biswas et al.29 15.1
Charitha and Manjunatha28 275
This work200

Electrochemical investigations allow sub-second detection of analytes of interest.44 The method also gives some insight into the electron transport process. To visualise the excitation of molecules under investigation, the redox current was plotted against the number of cycles and current in the form of two-dimensional hotspots (Figure 2(a)). This map allows tracking the electron-transfer kinetics of adrenaline. The plot shows an oxidation signature at 0.75 V and a second oxidation signature at 0.9 V. The reduction signal at 1.1 V (also in Figure 1(a)) has not been properly identified and is most likely the reduction of a side product of the reaction such as adrenochrome or an intermediate in the formation of peroxide and melanin. This observation is consistent with the authors’ previous recording of dopamine and serotonin on CNT and pyrolytic carbon surfaces (not shown), which shows similar behaviour on oxidative reaction. A reduction signature is seen at −0.2 V. From this, it is clear that adrenaline oxidation on CNT-pyrolytic electrodes is not a single step and consists of multiple intermediates.

To explain the electrochemical behaviour of adrenaline on CNT electrodes, the authors applied the ‘ECE’ model presented by Lin et al.45 and Bacil et al.46 A likely mechanism of adrenaline oxidation is the loss of an electron followed by rapid deprotonation. An intermediate radical is formed, which loses a further electron to form the quinone product.47Figure 2(b) shows a simulated voltammogram with these parameters compared with the measured data. While the peak current for experimental data is higher than those of the simulation, this is probably due to the shape of the break of the quartz capillaries increasing the effective area (Figures 2(c) and 2(d)). Both peaks are far broader than the simulated ones, and the reduction peak shows several maxima, suggesting a multistage oxidation and reduction process.

During the normal polymerisation of adrenaline (and other monoamines) to melanin, peroxides are formed.48 For this reason, it is useful to be able to measure peroxide and adrenaline simultaneously.

To evaluate if the pyrolytic-CNT electrode is capable of distinguishing analytes, the electrode was exposed to hydrogen peroxide and adenosine. Hydrogen peroxide is known to be oxidised at 1.4 V on glassy carbon electrodes.5,6 Background-subtracted voltammetry shows the oxidation of hydrogen peroxide on the surface of the CNT electrode (Figure 3(a)) where hydrogen peroxide is oxidised at 0.85 V and has no reduction peak. To understand the electrochemical kinetics of peroxide, the authors varied the scan rate and measured the peak current. The peak current was proportional to the scan rate, showing that peroxide is bound to the surface of the electrode (Figure 3(e)). The narrow oxidation peak at 0.85 V shows that oxidation of hydrogen peroxide is rapid and probably simple.5,6 Modelling the reaction shows the oxidation of hydrogen peroxide at 1.3 V, using the parameters from the study by Spanos et al.5 (Figure 3(b)). However, upon increasing the rate of electron transfer, the authors are able to match the theory to experimental data (Figure 3(c)). This agrees with previous work using oxidised CNTs27,28,31 and also doped diamond30 that electron-transfer rates from monoamines are higher at doped carbon than at pure carbon. Those studies used electrochemical impedance spectroscopy using a ferri/ferrocyanide couple to remove the effect of the overpotential.

Figure 3

Cyclic voltammogram of 200 μM hydrogen peroxide on a CNT-coated electrode. (a) Hydrogen peroxide oxidises at 0.85 V. (b) Electrochemical modelling of the reaction. The diffusion coefficient was 2.5 × 10−5 cm2/s, the charge-transfer coefficient was 0.5, the number of electron transfer (n) = 2 and the electrochemical rate was set to 10−4 cm/s. (c) Increasing the rate of reaction to 10−6 cm/s matches the experimental observation, thus suggesting that CNTs have a faster electron-transfer process. (d) Plot for 1 μM adenosine, which shows an oxidation peak at 0.1 V and a second oxidation at 1.1 V. Broad reduction peaks are seen at 0.7 and 0.2 V. (e) Plot of the scan rate against the peak current for hydrogen peroxide. The electrode has a sensitivity of 11 nA/μM

Figure 3

Cyclic voltammogram of 200 μM hydrogen peroxide on a CNT-coated electrode. (a) Hydrogen peroxide oxidises at 0.85 V. (b) Electrochemical modelling of the reaction. The diffusion coefficient was 2.5 × 10−5 cm2/s, the charge-transfer coefficient was 0.5, the number of electron transfer (n) = 2 and the electrochemical rate was set to 10−4 cm/s. (c) Increasing the rate of reaction to 10−6 cm/s matches the experimental observation, thus suggesting that CNTs have a faster electron-transfer process. (d) Plot for 1 μM adenosine, which shows an oxidation peak at 0.1 V and a second oxidation at 1.1 V. Broad reduction peaks are seen at 0.7 and 0.2 V. (e) Plot of the scan rate against the peak current for hydrogen peroxide. The electrode has a sensitivity of 11 nA/μM

Close modal

Adenosine interferes with the detection of hydrogen peroxide when measurements are conducted using normal carbon electrodes. To investigate the effect of adenosine, 1 μM of adenosine was introduced in the flow cell chamber. Adenosine shows a broad oxidation peak at 0.2 V and another at 1.2 V. Multiple reduction peaks can be observed at 0.6 and 0.2 V21 (Figure 3(d)). This suggests that pyrolytic carbon electrodes coated with CNTs have suitable overpotentials to separate the redox signals from analytes that have similar redox potentials on other substrates. Herein, the authors show that analytes with identical oxidation windows can be distinguished using CNT coatings onto pyrolytic electrodes. The authors’ future work will be setting up a calibration method for multimodal analysis of analytes for detecting the detection limit of their sensor. The authors also intend to work on the reproducibility of the processes, which currently require each electrode to be calibrated individually due to variations in tip diameter and possibly angle. One possibility is to wet or dry etch the tips of the capillaries before and possibly also after the pyrolysis process.

In this work, it was shown that oxidised CNT-coated pyrolytic carbon electrodes can be used for the detection of adrenaline, hydrogen peroxide and adenosine and give different signals for them. The ability of pyrolytic-oxidised CNT electrodes to distinguish between analytes by shifting the potential at which they are oxidised coupled with the low overpotential for monoamines (reducing potential oxidative damage from solvolysis) makes them potentially useful surfaces for the electrochemical detection of neuromodulators. As mentioned earlier, the use of modified carbon electrodes, particularly for this type of measurement, is established, but most have concentrated on relatively large electrodes,30,31 paste electrodes28,29 and some contain metal particles.29 

The authors’ future work will include determining the oxidation potential of adrenaline on different surfaces using the FSCV method and investigating ways to deconvolute mixed signals, as despite the voltage shift of the analytes, they are all adsorbed on the electrode and give broad signals, so they are not fully resolvable yet. The interaction between the substrate in solution and the electrode seems to be very complex, and the authors feel that understanding it more would enable the electrode surface to be modified to maximise the number of sites where adrenaline can interact or to generate greater electrochemical differences between similar substrates such as adrenaline and serotonin. It would also apply to similar systems, such as that of polyphenols.

1
Cechova
S
,
Venton
BJ
2008
Transient adenosine efflux in the rat caudate-putamen
Journal of Neurochemistry
105
4
1253
 -
1263
2
Jacobs
CB
,
Ivanov
IN
,
Nguyen
MD
,
Zestos
AG
,
Venton
BJ
2014
High temporal resolution measurements of dopamine with carbon nanotube yarn microelectrodes
Analytical Chemistry
86
12
5721
 -
5727
3
Hashemi
P
,
Dankoski
EC
,
Petrovic
J
,
Keithley
RB
,
Wightman
RM
2009
Voltammetric detection of 5-hydroxytryptamine release in the rat brain
Analytical Chemistry
81
22
9462
 -
9471
4
Swamy
BEK
,
Venton
BJ
2007
Carbon nanotube-modified microelectrodes for simultaneous detection of dopamine and serotonin in vivo
Analyst
132
9
876
 -
884
5
Spanos
M
,
Gras-Najjar
J
,
Letchworth
JM
, et al
2013
Quantitation of hydrogen peroxide fluctuations and their modulation of dopamine dynamics in the rat dorsal striatum using fast-scan cyclic voltammetry
ACS Chemical Neuroscience
4
5
782
 -
789
6
Meunier
CJ
,
Mitchell
EC
,
Roberts
JG
, et al
2018
Electrochemical selectivity achieved using a double voltammetric waveform and partial least squares regression: differentiating endogenous hydrogen peroxide fluctuations from shifts in pH
Analytical Chemistry
90
3
1767
 -
1776
7
Iijima
S
2002
Carbon nanotubes: past, present, and future
Physica B: Condensed Matter
323
1–4
1
 -
5
8
Ross
AE
,
Venton
BJ
2012
Nafion-CNT coated carbon-fiber microelectrodes for enhanced detection of adenosine
Analyst
137
13
3045
 -
3051
9
Yang
C
,
Trikantzopoulos
E
,
Jacobs
CB
,
Venton
BJ
2017
Evaluation of carbon nanotube fiber microelectrodes for neurotransmitter detection: correlation of electrochemical performance and surface properties
Analytica Chimica Acta
965
1
 -
8
10
Yang
C
,
Jacobs
CB
,
Nguyen
MD
, et al
2016
Carbon nanotubes grown on metal microelectrodes for the detection of dopamine
Analytical Chemistry
88
1
645
 -
652
11
Yang
C
,
Trikantzopoulos
E
,
Nguyen
MD
, et al
2016
Laser treated carbon nanotube yarn microelectrodes for rapid and sensitive detection of dopamine in vivo
ACS Sensors
1
5
508
 -
515
12
Schmidt
AC
,
Wang
X
,
Zhu
Y
,
Sombers
LA
2013
Carbon nanotube yarn electrodes for enhanced detection of neurotransmitter dynamics in live brain tissue
ACS Nano
7
9
7864
 -
7873
13
Kidokoro
Y
,
Ritchie
AK
1980
Chromaffin cell action potentials and their possible role in adrenaline secretion from rat adrenal medulla
Journal of Physiology
307
1
199
 -
216
14
Ambade
V
,
Arora
MM
,
Singh
P
,
Somani
BL
,
Basannar
D
2009
Adrenaline, noradrenaline and dopamine level estimation in depression: does it help?
Medical Journal Armed Forces India
65
3
216
 -
220
15
Lee
CR
,
Patel
JC
,
O’Neill
B
,
Rice
ME
2015
Inhibitory and excitatory neuromodulation by hydrogen peroxide: translating energetics to information
Journal of Physiology
593
16
3431
 -
3446
16
Rice
ME
2011
H2O2: a dynamic neuromodulator
Neuroscientist
17
4
389
 -
406
17
Dringen
R
,
Pawlowski
PG
,
Hirrlinger
J
2005
Peroxide detoxification by brain cells
Journal of Neuroscience Research
79
1–2
157
 -
165
18
Petrovic
J
,
Walsh
PL
,
Thornley
KT
,
Miller
CE
,
Wightman
RM
2010
Real-time monitoring of chemical transmission in slices of the murine adrenal gland
Endocrinology
151
4
1773
 -
1783
19
Lee
ST
,
Venton
BJ
2018
Regional variations of spontaneous, transient adenosine release in brain slices
ACS Chemical Neuroscience
9
3
505
 -
513
20
Ross
AE
,
Venton
BJ
2014
Sawhorse waveform voltammetry for selective detection of adenosine, ATP, and hydrogen peroxide
Analytical Chemistry
86
15
7486
 -
7493
21
Heien
MLAV
,
Phillips
PEM
,
Stuber
GD
,
Seipel
AT
,
Wightman
RM
2003
Overoxidation of carbon-fiber microelectrodes enhances dopamine adsorption and increases sensitivity
Analyst
128
12
1413
 -
1419
22
Weese
ME
,
Krevh
RA
,
Li
Y
,
Alvarez
NT
,
Ross
AE
2019
Defect sites modulate fouling resistance on carbon-nanotube fiber electrodes
ACS Sensors
4
4
1001
 -
1007
23
Trautner
E
,
Bradley
T
1951
The early stages of the oxidation of adrenaline in dilute solution
Australian Journal of Scientific Research Series B
4
3
303
 -
343
24
Chirila
T
,
Cooper
R
,
Constable
I
1991
‘Black Prosthesis’ revisited: a study of epinephrine-induced pigment deposits on poly(methyl methacrylate)
Graefe’s Archive for Clinical and Experimental Ophthalmology
229
6
578
 -
582
25
Swamy
BEK
,
Venton
BJ
2007
Subsecond detection of physiological adenosine concentrations using fast-scan cyclic voltammetry
Analytical Chemistry
79
2
744
 -
750
26
Nasirizadeh
N
,
Shekari
Z
,
Zare
H
, et al
2013
Electrosynthesis of an imidazole derivative and its application as a bifunctional electrocatalyst for simultaneous determination of ascorbic acid, adrenaline, acetaminophen, and tryptophan at a multi-wall carbon nanotubes modified electrode surface
Biosensors and Bioelectronics
41
608
 -
614
27
Reddy
KK
,
Satyanarayana
M
,
Goud
KY
,
Gobi
KV
,
Kim
H
2017
Carbon nanotube ensembled hybrid nanocomposite electrode for direct electrochemical detection of epinephrine in pharmaceutical tablets and urine
Materials Science and Engineering: C
79
93
 -
99
28
Charithra
MM
,
Manjunatha
JG
2021
Electrochemical sensing of adrenaline using surface modified carbon nanotube paste electrode
Materials Chemistry and Physics
262
article 124293
29
Biswas
S
,
Naskar
H
,
Pradhan
S
, et al
2017
Simultaneous voltammetric determination of adrenaline and tyrosine in real samples by neodymium oxide nanoparticles grafted graphene
Talanta
206
article 120176
30
Li
YR
,
Li
HJ
,
Li
MJ
,
et
al.
2017
Porous boron-doped diamond electrode for detection of dopamine and pyridoxine in human serum
Electrochimica Acta
258
744
 -
753
31
Zare
HR
,
Moradiyan
B
,
Shekari
Z
,
Benvidi
A
2016
Application of l-DOPA modified carbon nanotubes as a bifunctional electrocatalyst for simultaneous determination of ascorbic acid, adrenaline, acetaminophen and tyrosine
Measurement
90
510
 -
518
32
Liu
S
,
Ko
A
,
Li
W
,
Zhong
W
,
Xing
M
2014
NIR initiated and pH sensitive single-wall carbon nanotubes for doxorubicin intracellular delivery
Journal of Materials Chemistry B
2
9
1125
 -
1135
33
Kole
K
2015
Experience dependent plasticity of neurovascularisation
Journal of Neurophysiology
114
4
2077
 -
2079
34
da Silva Lantyer
A
,
Calcini
N
,
Bijlsma
A
, et al
2018
A databank for intracellular electrophysiological mapping of the adult somatosensory cortex
Gigascience
7
12
article giy147
35
Kole
K
,
Zhang
Y
,
Jansen
EJR
, et al
2020
Assessing the utility of Magneto to control neuronal excitability in the somatosensory cortex
Nature Neuroscience
23
9
1044
 -
1046
36
Kole
K
,
Celikel
T
2019
Neocortical microdissection at columnar and laminar resolution for molecular interrogation
Current Protocols in Neuroscience
86
1
article e55
37
Li
S
,
Wang
H
,
Young
M
, et al
2018
Properties of electropolymerized dopamine and its analogues
Langmuir
35
5
1119
 -
1125
38
Kund
J
,
Daboss
S
,
D’Alvise
T
, et al
2021
Physicochemical and electrochemical characterization of electropolymerized polydopamine films: influence of the deposition process
Nanomaterials
11
8
article 1964
39
Thirumalraj
B
,
Palanisamy
S
,
Chen
S
,
Kannan
R
2016
Alumina polished glassy carbon electrode as a simple electrode for lower potential electrochemical detection of dopamine in its sub-micromolar level
Electroanalysis
28
3
425
 -
430
40
Kole
K
,
Scheenen
W
,
Tiesinga
P
,
Celikel
T
2018
Cellular diversity of the somatosensory cortical map plasticity
Neuroscience & Biobehavioral Reviews
84
100
 -
115
41
Atcherley
CW
,
Laude
ND
,
Parent
KL
,
Heien
ML
2013
Fast-scan controlled-adsorption voltammetry for the quantification of absolute concentrations and adsorption dynamics
Langmuir
29
48
14885
 -
14892
42
Yuan
Z
2014
The effects of abrasive particles on the electrochemical behavior of adrenaline at different electrodes
Russian Journal of Electrochemistry
50
1
80
 -
86
43
Yi
Y
,
Weinberg
G
,
Prenzel
M
, et al
2017
Electrochemical corrosion of a glassy carbon electrode
Catalysis Today
295
32
 -
40
44
Venton
BJ
,
Troyer
KP
,
Wightman
RM
2002
Response times of carbon fiber microelectrodes to dynamic changes in catecholamine concentration
Analytical Chemistry
74
3
539
 -
546
45
Lin
C
,
Chen
L
,
Tanner
EEL
,
Compton
RG
2017
Electroanalytical study of dopamine oxidation on carbon electrodes: from the macro- to the micro-scale
Physical Chemistry Chemical Physics
20
1
148
 -
157
46
Bacil
RP
,
Chen
L
,
Serrano
SHP
,
Compton
RG
2020
Dopamine oxidation at gold electrodes: mechanism and kinetics near neutral pH
Physical Chemistry Chemical Physics
22
2
607
 -
614
47
Choi
J
,
Tojo
S
,
Ahn
DS
, et al
2019
Proton transfer accompanied by the oxidation of adenosine
Chemistry – A European Journal
25
32
7711
 -
7718
48
Schales
O
1938
Empfindliche Methoden zum Nachweis von Wasserstoffperoxyd und seiner Entstehung bei einigen Dehydrierungs-Vorgängen
Berichte der Deutschen Chemischen Gesellschaft
71
2
447
 -
460 (in German)
This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

or Create an Account

Close Modal
Close Modal