Purpose

The purpose of this paper is to evaluate the dosimetric properties of N-(3-methoxypropyl) acrylamide (NMPA) polymer gel dosimeter using UV-vis spectrophotometry as a simple and low-cost method.

Design/methodology/approach

The gel samples have been irradiated to various doses using a medical linear accelerator with 6 MV beam energy. The optical density of un-irradiated and irradiated NMPA polymer gel dosimeters in terms of absorbance at 500 nm was investigated by UV-vis spectrophotometry.

Findings

The absorbance values of the gel samples were increased linearly with increasing dose in the dose range between 2 and 20 Gy. A remarkable improvement in the dose response was noticed after the addition of various concentrations of glycerol. The impact of dose rate, beam energy and the post-stability of the exposed gels was studied and found consistent with the previous study using nuclear magnetic resonance results (energy independent and dose rate independent, stable up to one week).

Practical implications

The results of the independent experimental spectrophotometry and nuclear magnetic resonance analyses indicate that the NMPA polymer gel dosimeter has good and consistent dosimetric features for radiotherapy dosimetry.

Originality/value

The ionizing radiation-induced polymerization of gel samples leads to a change in the absorbance of the irradiated gel samples. This study introduces the first characterization of NMPA gel dosimeter by means of UV-vis spectrophotometer.

Licensed re-use rights only
You do not currently have access to this content.
Don't already have an account? Register

Purchased this content as a guest? Enter your email address to restore access.

Please enter valid email address.
Email address must be 94 characters or fewer.
Pay-Per-View Access
$41.00
Rental

or Create an Account

Close Modal
Close Modal

Gift article access

As a benefit of your subscription, you can share temporary access to restricted articles.

Each link will stop working after 30 days or 10 uses. You may create up to 10 links in a 30 day period.

Please sign in to your personal account to gift article access.

Register

Gift article access

As a benefit of your subscription, you can share temporary access to restricted articles.

Each link will stop working after 30 days or 10 uses. You may create up to 10 links in a 30 day period.

Gift articles remaining: --

Gift article access

Each link will stop working after 30 days or 10 uses. You may create up to 10 links in a 30 day period.

Gift articles remaining: --

Gift article access

As a benefit of your subscription, you can share temporary access to restricted articles.

Each link will stop working after 30 days or 10 uses.

You have reached the limit of 10 links within a 30 day period.