An important challenge of the osteotomy procedures, particularly in the case of large and complex corrections, is the fixation of the osteotomy site. The purpose of this study is to propose a practical and cost-effect methodology for the plate adapting problem of osteotomy surgery.
A novel patient-specific plate contouring methodology, based on rapid prototyping (RP) and multi-point forming (MPF) techniques, was developed and evaluated. In this methodology, a female mold is fabricated by RP, based on the geometry of the osteotomy site and estimation of the plate spring back. The mold is then used to configure a MPF die, which is then used for press forming of the factory-made locking plate. The applicability of the methodology was assessed in two case studies.
The results of implementing the methodology on a femoral and a tibial locking plate indicated very good conformity with the underlying bone, in both the frontal and sagittal planes. The surgical application of the pre-operatively contoured tibial plate facilitated the plate locating and screw inserting procedures, and provided a secure fixation for bone fragments.
The results are promising and provide a proof of concept for the feasibility and applicability of the proposed methodology in clinical practice, as a complementary to the existing surgical preplanning and patient-specific instrument preparations.
The advantageous features of RP and the MPF were used to provide a solution for the plate adapting problem of osteotomy surgery.