Article navigation
Purpose

Three-dimensional (3D) printing is increasingly used to produce orthopaedic components for hip arthroplasty, such as acetabular cups, which show complex lattice porous structures and shapes. However, limitations on the quality of the final implants are present; thus, investigations are needed to ensure adequate quality and patients safety. X-ray microcomputed tomography (micro-CT) has been recognised to be the most suitable method to evaluate the complexity of 3D-printed parts. The purpose of this study was to assess the reliability of a micro-CT analysis method comparing it with reference systems, such as coordinate measuring machine and electron microscopy.

Design/methodology/approach

3D-printed acetabular components for hip arthroplasty (n = 2) were investigated. Dimensions related to the dense and porous regions of the samples were measured. The micro-CT scanning parameters (voltage – kV, current – µA) were optimised selecting six combinations of beam voltage and current.

Findings

Micro-CT showed good correlation and agreement with both coordinate measuring machine and scanning electron microscopy when optimal scanning parameters were selected (130 kV – 100 µA to 180 kV – 80 µA). Mean discrepancies of 50 µm (± 300) and 20 µm (± 60) were found between the techniques for dense and porous dimensions. Investigation method such as micro-CT imaging may help to better understand the impact of 3D printing manufacturing technology on the properties of orthopaedic implants.

Originality/value

The optimisation of the scanning parameters and the validation of this method with reference techniques may guide further analysis of similar orthopaedic components.

Licensed re-use rights only
You do not currently have access to this content.
Don't already have an account? Register

Purchased this content as a guest? Enter your email address to restore access.

Pay-Per-View Access
$41.00
Rental

or Create an Account

Close subscription notice
Close access options