The purpose of this paper is to show how a quartz crystal microbalance modified with poly (epichlorohydrin) (PECH) is fabricated and used for detection of acetone vapor using a spin‐coating method.
Measurements are based on the frequency shifts due to the adsorption of acetone vapor on the surface of the modified electrodes. A first‐order linear time‐invariant (LTI) system model is used to simulate the sensor response.
It is found that frequency shifts are linear to the concentrations of analyte, and cycle test examination shows good repeatability. The sensitivity to acetone vapor is better than other volatile compounds. A good match result is obtained with the LTI system model.
The device performance opens a potential way for analyzing acetone vapor.